VF-180 MICROELECTRODE AMPLIFIER


 

The VF-180 is a high input impedance voltage follower for intracellular voltage measurements. The VF-180 has a current injection capability and possesses a true current generator which adjusts the injected current so that it is always proportional to the command pulse amplitude (1 nA for 100 mV) and independent of the charge resistance.

VF-180
Click
to enlarge

Standard head-stage is HS-180. Three other head-stages are available for the VF-180. Choice is dependent upon current injection requirements. Head-stages can be interchanged without internal recalibration of the instrument. Each head-stage is fitted with a miniature, coaxial electrode holder with driver shield. Only 5 mm in diameter, gold plated and 50 mm long, the electrode holder is a masterpiece in design.

FEATURES
  • High impedance head-stage
  • True current generator
  • Negative capacity compensation
  • Oscillation burst (tickle) facility
  • Electrode resistance compensation (Bridge)
SPECIFICATIONS

Rise time (HS-180)

  • 0.5 us with pulse applied directly to the input
  • 20 us on 10 MOhm load fully compensated for input capacitance

Input capacity

  • Around 1 pF, adjusted to zero by negative capacity compensation

Frequency response

  • 2 MHz on x 1 setting - 3 dB at 500 kHz on x 10 setting
  • Low pass filter : - 3 dB at 1 or 10 kHz

Noise spectra

  • Noise measured after amplification (x 10 gain setting with HS-180 probe)
  • Zero source resistance : 10 µV RMS with 1 kHz filter
  • 10 MOhm source resistance : 30 µV RMS with 1 kHz filter

Notch-filter

  • 50 or 60 Hz Notch-filter

Negative capacity compensation

  • 0 to 33 pF

Voltage measurements

  • Gain : x 1, x 10. Junction null : +/- 200 mV

Electrode resistance compensation (Bridge)

  • Two scales on a ten-turn dial. 100 MOhm/1000 MOhm with HS-180

Current injection

  • Input for current command (1000 mV = 1 nA with HS-180)
  • Maximum leakage current : 300 fA adjustable to zero
  • Calibrated 1 nA injection for resistance measurements

Power supply

  • 220 V/50 Hz, 110 V/60Hz

Head-stages

Head-Stage H-160 H-170 H-180 H-111
Input impedance (Ohm) >1E9 >1E10 >1E11 >1E14
Recommended for electrode resistance 1-10MOhm 10-100MOhm 100MOhm Ion selective electrodes
Bias current <300 fA <300 fA <300 fA <300 fA
Maximum current injection 10 µA 1.0 µA 100 nA  

 

ORDERING INFORMATION

The VF-180 is delivered complete with one HS-180 head-stage (ref. 013-08) and with one SM-150 electrode-holder (for electrodes of diameter 1.5mm). Please specify on order if you want to receive other types of head-stage or electrode-holder.

 